AS/NZS 2243.8: Fume Cupboard Standards

AS/NZS 2243.8 covers safety requirements for fume cupboards and their selection, installation, testing and use. Start by identifying the edition and requirements applicable to your project, then connect the standard with the cupboard, room and extraction system.

At a glance

  1. 01Confirm the edition used by the project.
  2. 02Separate cupboard performance from room design.
  3. 03Keep the specified tests and records together.

What AS/NZS 2243.8 covers

MBIE's CodeHub entry for AS/NZS 2243.8:2014 describes its scope as safety requirements for fume cupboards, including performance, selection, installation, testing and use. It also identifies additional requirements for special applications in Appendix D.

The standard is a technical reference that must be read in full for the relevant project. This overview helps you prepare questions; it does not replace the standard, a project specification or advice from the responsible designer and compliance professionals.

Record the edition in the design brief and test scope. A standard's catalogue status and the edition cited by a particular requirement are different questions. Confirm both through the relevant official documents.

AS/NZS 2243.8 and recirculating cupboards

The CodeHub scope states that recirculating fume cabinets are not covered by AS/NZS 2243.8 and refers to AS/NZS 2243.9. Establish whether the proposed equipment exhausts outside or recirculates filtered air before choosing the assessment basis.

That distinction affects the system and the information required about the chemical process. Our ducted and ductless fume hood comparison explains the equipment choice. A biological safety cabinet also has a different protection purpose; it should not be treated as interchangeable with a chemical fume cupboard.

Selection and installation questions

Begin with the chemicals, quantities, temperatures and apparatus. Include any special process such as heated work or distillation. The required working envelope, services and material compatibility should follow that task.

  • Which cupboard configuration and working opening are specified?
  • How will the duct, fan and discharge arrangement achieve the required duty?
  • How will room supply air and nearby movement affect the opening?
  • What alarms, controls and services are required?
  • Which drawings, manuals and test records will be handed over?

Our installation planning guide helps coordinate the building and equipment responsibilities. Keep changes to apparatus or room layout within that review.

Testing and a useful report

Agree the test scope, operating conditions and acceptance basis before testing. Record the cupboard identification, sash position, room conditions, instruments, results and any limitations. A report needs enough context for a later reviewer to understand what was tested.

Different methods answer different questions. Airflow measurements, visualisation and containment tests are not interchangeable labels for one result. Our ASHRAE 110 guide and EN 14175 guide explain other test frameworks without treating them as automatic substitutes for a project's specified requirements.

For New Zealand building-related inspection and reporting, discuss the applicable schedule and responsibilities with the relevant professionals. Our SS11 fume cupboard testing service is the enquiry route for that work.

Keep the standard connected to everyday use

Retain the operating instructions, test reports and maintenance record with the equipment. Make the intended working opening and permitted use clear to laboratory users. Changes in chemicals, apparatus, room ventilation or controls can require a review of the original basis.

Use the fume hood safety guide to organise user information and routine checks. A successful handover should explain who reports a fault, who can alter settings and what happens when the extraction is unavailable.

Your questions

Is AS/NZS 2243.8 a universal certificate for every cupboard?

No. A claim or report needs to identify the equipment, edition, assessment scope and results. Confirm what the project requires and what the supplied documentation actually covers.

Does it cover recirculating fume cabinets?

The published scope for AS/NZS 2243.8:2014 excludes recirculating fume cabinets and refers to AS/NZS 2243.9. Confirm the applicable reference for the proposed equipment.

Bring the project requirements together

Send the cupboard details, chemical task and specified testing or reporting scope. We can discuss selection, installation and the FumeCare testing enquiry that fits your project.